Replacement E047 filament assembly, part number 129108S, for a compatible 3G helium leak-detector analyzer cell. Compatibility must be verified from the instrument model, serial number and service documentation before installation.
Description
E047 Filament, Part 129108S, for 3G Analyzer Cell
This service component contains an iridium filament assembly for a compatible helium mass-spectrometer analyzer cell. The source record lists stainless steel, polymer and iridium among its materials.
Do not select an analyzer filament by appearance alone. Confirm detector manufacturer, exact model, analyzer-cell revision, part number 129108S and instrument serial-number range. Installation must follow the service manual's clean-handling, torque, alignment, electrical and vacuum-conditioning procedure.
After replacement, perform the prescribed analyzer checkout, tuning, background stabilization, reference-leak response test and any required calibration. Opening the analyzer can affect warranty, cleanliness and high-voltage safety.
Specifications
| Product Label | New |