Heated absolute-pressure capacitance diaphragm gauge with a remote-zero input and particle-trap feature. DM1 configurations cover selected full scales from 0.1 to 1000 Torr, with 45 °C or 100 °C temperature control and a linear 0–10 V output.
Description
CBVAC DM1 Capacitance Diaphragm Gauge
The DM1 adds remote-zero control and a particle-control feature to a heated capacitance diaphragm pressure transducer. It measures absolute pressure independently of gas thermal conductivity; process compatibility remains material- and condition-specific.
| Available full-scale ranges stated in source record | 0.1, 0.25, 1, 2, 10, 20, 100 and 1000 Torr |
|---|---|
| Accuracy stated in source record | ±0.12% of reading for applicable 1–1000 Torr versions; ±0.15% for 0.1 and 0.25 Torr versions |
| Resolution | 0.001% of full scale |
| Response time | <20 ms |
| Heater setpoint options | 45 °C or 100 °C |
| Analog output | 0–10 V DC; load ≥10 kΩ |
| Supply | ±15 V DC ±5%; current depends on heater version |
| Electrical connector | DB9 male with remote-zero input in the source configuration |
| Process connections | KF16, KF25, CF16, CF35, 8VCR, 8VCO and 1/2-inch tube options stated in source record |
| Maximum overpressure stated in source record | 310 kPa |
| Wetted material stated in source record | Corrosion-resistant nickel alloy |
Remote zero must be used only under the pressure and warm-up conditions specified by the manufacturer; zeroing at an unsuitable pressure hides measurement error. The internal particle trap does not replace upstream process protection.
Confirm full scale, heater, fitting, remote-zero logic, pinout, wetted materials, accuracy basis and calibration before ordering.